Email us : sales@ddmeter.com
|
Categories
Homebrew
DS-111 500g/0.01g Electronic Balance Diamond Electronic Pocket Laboratory Balance Scale Jewelry Scale
  • DS-111 500g/0.01g Electronic Balance Diamond Electronic Pocket Laboratory Balance Scale Jewelry Scale
  • DS-111 500g/0.01g Electronic Balance Diamond Electronic Pocket Laboratory Balance Scale Jewelry Scale
  • DS-111 500g/0.01g Electronic Balance Diamond Electronic Pocket Laboratory Balance Scale Jewelry Scale
  • DS-111 500g/0.01g Electronic Balance Diamond Electronic Pocket Laboratory Balance Scale Jewelry Scale
  • DS-111 500g/0.01g Electronic Balance Diamond Electronic Pocket Laboratory Balance Scale Jewelry Scale
  • DS-111 500g/0.01g Electronic Balance Diamond Electronic Pocket Laboratory Balance Scale Jewelry Scale

DS-111 500g/0.01g Electronic Balance Diamond Electronic Pocket Laboratory Balance Scale Jewelry Scale

Item No.: DS-111
INQUIRY

I. Parameter:
Double precision: 0-100g/0.01g-101-500g/0.1
Operating temperature: 10-30 ℃ Automatic shutdown: 1 minute Automatic shutdown (adjustable)
LCD display: 5-digit LCD display backlight: white backlight (optional)
Power supply: CR2032 x 1 Appearance size: 40X75X15.5mm
Unit weight: 38g Stainless steel tray: 36.2x41mm
Low point alarm LO for automatic low point function
Unit: g/oz/T/gn/ozt/dwt/ct II.
Operating instructions:
1. Place the scale body on a stable platform Press the [ON] key and wait for 3-5 seconds for "0.0" to appear
2. Place the heavy object on the tray surface, and when the number stabilizes, it becomes the weight of the object being weighed
3. Use the [MODE] key to change the measurement unit: g/oz/T/gn/ozt/dwt/ct
4. Press the [TARE] key to force zero
The ON/OFF button is used to turn on and off the device
Mini Pocket Scale Instructions (DS-111) III.
Calibration: (Please use this function if weighing is not accurate due to large temperature differences or other reasons)
1. Press the [ON/OFF] button to turn on the machine  
2. After zeroing, press the ON button again for 3 seconds until the screen shows no display, wait for another 3 seconds until "0" is displayed, release the button, and display the verification value (such as 500.0100.00),Unit flashing
3. (Pressing TARE or MODE can change the weight value)  At this point, place the corresponding weight in the center of the tray. After about 5 seconds, the unit will stop flashing and display the current weight of the weight  
4. Calibration completed! IV. Attention: Gently remove the tray protective cover and place the scale body in a stable and vibration free environment for use *Please do not place any objects larger than the rated weight on the tray, as this may damage the sensor. When placing objects, do not impact the weighing tray and handle them with care  
*Do not use mobile phones within 50cm around the scale body when using this product, as this will affect the use of this product *When the battery symbol is displayed on the screen, please replace the battery immediately  
*Please use this product in an environment of 10-30 ℃, as a large temperature difference may cause testing errors If the temperature difference is too large, please use standard weights to calibrate the original product before use.
During calibration, it is necessary to ensure that this product remains in the same environment for at least 30 minutes *When the weight added to the tray exceeds the rated weight, the "FULL" symbol will appear